CRY3103 Free-field Microphone, 1", Prepolarized, 100 mV/Pa

CRY3103 is a 1 inch free-field prepolarized measurement microphone, widely applicable in acoustic test environments.
The unique design of CRY3103 endows it with high sensitivity, providing users with excellent test performance and flexibility. It is widely applicable in acoustic testing, acoustic design research, and calibration testing of audio equipment such as speakers.

Downloads
Product Highlights
CRY3103 Measurement Microphone Typical Frequency Response

CRY3103 Measurement Microphone Typical Frequency Response

The graph (right) shows the frequency response curves for 0° incidence in a free-field environment and the frequency response curves obtained using the electrostatic excitation method.

CRY3103 Measurement Microphone Typical Frequency Response
Technical Specifications
Field Type
Free-field
Sensitivity
100 mV/Pa, -20 dB±2 dB re 1 V/Pa
Frequency Response
4 Hz to 12 kHz ±2 dB
Polarization Voltage
0 V
Capacitance
66 pF (@250 Hz)
Dynamic Range(re.20uPa)
10 dB to 130 dB
Inherent noise
10 dBA
Operating Temperature
-20°C to +60°C(-4℉ to +140℉)
Temperature Stability
0.015 dB/℃ (-10℃ to +50℃) ; 0.008 dB/℉ (+14℉ to +122℉)
Static Pressure Stability
-0.01 dB/kPa
Operating Humidity Range
0 to 90%RH no condensation
Humidity Stability
< 0.1 dB (0 to 90%RH no condensation)
Pressure Equalization Vent
Rear vented
IEC 61094-4 Type
WS1F
Height with Grid
19 mm (0.748")
Diameter with Grid
23.77 mm (0.936")
Main Features

Related Products

CRYSOUND 10-Year Warranty for CRY3000 Series Sensors

Precision measurement is only as trustworthy as the tools behind it. At CRYSOUND, long-term reliability has always been part of how we support professional acoustic testing and metrology work-especially for equipment expected to perform consistently over many years. That's why CRYSOUND provides a 10-Year Limited Warranty for eligibleCRY3000 series sensors offering long-horizon confidence for labs, [...]

Microphone Sound Fields: Free, Pressure & Diffuse Guide

In acoustic measurements (SPL, frequency response, noise, reverberation, etc.), large errors often come not from instrument accuracy, but from a mismatch between the assumed sound field and the actual one. What a microphone reads as sound pressure is not strictly equivalent across different fields-especially at mid and high frequencies, where the microphone dimensions become comparable [...]

Microphone Test System

CRYSOUND provides a microphone test system test system solution.

How a 10-Year Warranty Lowers Measurement Microphone Costs

As the core component of measurement chains, the long-term stability of measuring microphones directly impacts the comparability and traceability of measurement data. The 10-year limited warranty (hereinafter referred to as the 10-year warranty) is not merely a service commitment, but a comprehensive capability demonstrated through manufacturing consistency control, reliability verification systems, and traceable evidence chains. [...]

Guide to Common Measurement Microphone Interfaces

From the outside, a measurement microphone looks deceptively simple. But in real-world engineering, its interface options are surprisingly diverse: Lemo, BNC, Microdot, 10-32 UNF, M5, SMB... Many newcomers to acoustics ask questions like: This article provides a structured overview of common measurement microphone interfaces, looking at physical connectors, powering methods, cable characteristics, and typical application-driven [...]

Get in touch

If you are interested or have questions about our products, book a demo and we will be glad to show how it works, which solutions it can take part of and discuss how it might fit your needs and organization.

Downloads Downloads
Quote 0
Quote
+86-571-88225128 +86-571-88225128
Experience Demo Experience Demo
Ask an Expert Ask an Expert