CRY3711 Ear Simulator IEC60318-4
The CRY3711 ear simulator simulates the way an earplug conduit is inserted into the ear canal or auricle to measure headphone performance. It has a 1/2 inch prepolarized pressure field measurement microphone inside and can be used for acoustic testing of high-quality in-ear headphones.
The input impedance of CRY3711 is very close to the input impedance of an ordinary human ear, enabling effective measurement up to 10 kHz.
Product Highlights
CRY3711 Ear Simulator Typical Frequency Response
The graph (right) shows the frequency response curves obtained using the electrostatic excitation method.
Technical Specifications
Nominal Sensitivity
12.5mV/Pa, -38±1.5dB re 1V/Pa
Frequency Response
100Hz to 10kHz ±1dB(Simulated human ear impedance)、 20Hzto 16kHz(Coupling cavity use)
Dynamic Range
23dB to 160dB
Standards
IEC 60318-4,ITU-T P.57 Type 2, SJ/T 10659-1995
Equivalent Volume
1.26cc
Weight
72g
Main Features
- IEC 60318-4 Electroacoustics-Simulators of human head and ear-Prat 4
- Extensive compatibility
- Durable, reliable, and high-quality
Related Products
CRY3718 Ear Simulator IEC60318-1
Frequency Response
100Hz to 4kHz ±1dB(Simulated human ear impedance)、 20Hz to 16kHz(Coupling cavity use)
Standards
IEC 60318-1、 ITU-T P.57 Type 1、 GB/T 25498.1-2010
Weight
147g
CRY3719 Coupler IEC 60318-5
Frequency Response
125Hz to 8kHz
Equivalent Volume
2cc
Standards
IEC60318-5, ANSI S3.7、 GB/T 20441.4-2006
CRY3717 IEC 60318-3 Coupler
Frequency Response
100Hz to 8kHz±2dB
Standards
IEC60318-3
Equivalent Volume
6cc
CRY3721 Ear Simulator IEC60318-4, Low-Noise
Standards
IEC 60318-4, ITU-T P.57 Type 2
Equivalent Volume
1.26cc
Weight
50g